ClearanceBrief
Recall Z-1139-2023 · CBK Ventilator, Continuous, Facility Use

Baxter Healthcare Corporation

Life2000 Ventilator System, a) REF BT-20-0002, b) REF BT-20-0002A, ventilatory support for the care of individuals

DateFeb 24, 2023 (initiated Jan 25, 2023)
StatusOpen, Classified
ReasonThere is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Root causeUnder Investigation by firm
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its healthcare providers, wholesalers, distributor/reseller, and durable medical equipment providers on 01/25/2023 by USPS First Class Mail. The notice explained the issue and the hazard and requested the following actions be taken by adressee: "Please follow the daily checks and preventive maintenance requirements below to ensure the best oxygen delivery with the Life2000 ventilation system when used with a third-party oxygen concentrator. 1- Daily Checks a) Check the CombO2 hose and Breathe Pillow Interface for kinks, dirt, moistu
Quantity4279 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170037Breathe Technologies Life2000™ Ventilation System
Breathe Technologies, Aug 11, 2017