Recall Z-1046-2023 ยท KRR Prosthesis, Knee, Patello/Femoral, Semi-Constrained, Cemented, Metal/Polymer

MEDLINE INDUSTRIES, LP - Northfield

MEDLINE UNITE FOOT & ANKLE Screw, Polyaxial Locking, Mini T8 (2.0x8mm), REF MMSL2008

DateFeb 1, 2023 (initiated Dec 22, 2022)
StatusOpen, Classified
ReasonThe recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.
Root causeDevice Design
ActionMedline issued a MEDICAL DEVICE RECALL notice to its direct consignees on 01/05/2023 by US mail and its distributors on 01/07/2023 by email. The notice explained the problem and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected items. Quarantine all affected product. All consignment inventory will be addressed by your local Medline UNITE Sales Representative. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Even if you do not have any affected product please complete and return the form, as indi
Quantity420 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181280Patello-Femoral Wave (Kahuna) Arthroplasty System
Arthrosurface, Inc., Jun 14, 2018