Recall Z-1042-2023 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Biomet, Inc.

(1) Vanguard Knee System PS Open Box Femoral; Right; 60 mm. Intended for knee joint replacement prostheses. Item Number: 183104 (2) Vanguard Knee System PS Open Box Femoral; Right; 62.5 mm Item Number: 183106 (3) Vanguard Knee System PS Open Box Femoral; Right; 65 mm Item Number: 183108 (4) Vanguard Knee System PS Open Box Femoral; Right; 67.5 mm Item Number: 183110 (5) Vanguard Knee System

DateJan 30, 2023 (initiated Dec 29, 2022)
StatusOpen, Classified
ReasonComponents Incorrectly labeled as either smaller or larger and incorrect side (right/left).
Root causeLabeling Change Control
ActionUPDATE: 2/7/23 Zimmer Biomet issued URGENT MEDICAL DEVICE RECALL EXPANSION Letter on 2/07/23 to Distributors, Risk Manager, Surgeons. (An Urgent Medical Device Letter issued to consignees new accounts not previously notified Dec 29, 2022). Letters state reason for recall, health risk and action to take: This is a 14-lot scope expansion to the recall notification initiated on December 29, 2022, for the Vanguard Posterior Stabilized Open Box Femoral components. Letter states reason for recall, health risk and action to take: . Immediately locate and quarantine affected product in your invento
Quantity157 units
DistributionInternational distribution to the countries of: Argentina, China, Costa Rica, India, Japan, Mexico, Netherlands, Ecuador, Republic of Korea, Singapore, and Thailand. Expanded Recall US states of: FL, IN, MD, OH, PA, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113550VANGUARD COMPLETE KNEE SYSTEM
Biomet Manufacturing Corp, Feb 29, 2012