Recall Z-1200-2023 ยท KDD Kit, Surgical Instrument, Disposable

Beaver Visitec International, Inc.

CustomEyes kits

DateMar 1, 2023 (initiated Dec 20, 2022)
StatusOpen, Classified
ReasonDevice packaging may contain open seals, compromising product sterility.
Root causeProcess control
ActionA consignee letter was issued to impacted customers on January 18, 2023. The communication advises consignees to immediately examine inventory and quarantine product from impacted lots. If further distribution has occurred, a sub-recall should be initiated. A completed copy of the Business Response Form shall be completed and returned to beavervisitec8978@sedgwick.com.
Quantity4640 units in total
DistributionUS distribution in AL, CA, CO, DC, FL, GA, IL, LA, MD, MN, MT, NC, NH, NJ, NM, NY, OH OK, PA, RI, TN, TX, UT, VA, VT, WA, and WI OUS distribution to Canada, Japan, and South Africa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.