Recall Z-1096-2023 ยท OJH Orthopedic Tray

MEDLINE INDUSTRIES, LP - Northfield

Sterile Procedural Trays, labeled as the following: a. SPINAL b. SPINE CDS c. TOTAL ORTHO CDS d. TOTAL KNEE CDS e. CDS HIP-LF f. KNEE TOTAL PACK-LF g. EXTREMITY CDS h. SHOULDER ARTHRO CDS i. GLENNON EXTREMITY CDS j. KNEE ARTHROSCOPY k. TOTAL KNEE l. TOTAL HIP REPLACEMENT m. TOTAL KNEE CDS-LF n. TOTAL SHOULDER CDS-LF o. TOTAL KNEE CDS-3 p. TOTAL HIP CDS-3 q. LOWER EXTREMITY CDS-LF r. TOTAL HIP-L

DateFeb 6, 2023 (initiated Nov 17, 2022)
StatusOpen, Classified
ReasonSterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Root causeDevice Design
ActionMedline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Quantity101,102 cases (227,538 units)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.