ClearanceBrief
Recall Z-1095-2023 ยท OJK Eye Tray

MEDLINE INDUSTRIES, LP - Northfield

Sterile Procedural Trays, labeled as the following: a. BASIC EYE PACK-LF b. VITRECTOMY PACK-LF c. EYE PACK-EAST-LF d. OSC OCULAR PLASTIC PACK-LF e. EYE PACK f. OCULOPLASTIC PACK g. EYE PLASTIC PACK h. PLASTIC EYE PACK i. DR H BLEPH PACK j. BLEPH PACK k. OPHTHALMOLOGY PACK l. OCULOPLASTIC PACK m. BLEPH PACK n. EYE PACK o. OCULAR PLASTIC PACK p. OCULOPLASTIC PACK

DateFeb 6, 2023 (initiated Nov 17, 2022)
StatusOpen, Classified
ReasonSterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Root causeDevice Design
ActionMedline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Quantity260 cases (1,200 units)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.