ClearanceBrief
Recall Z-1091-2023 ยท OHD Gynecological Laparoscopic Kit

MEDLINE INDUSTRIES, LP - Northfield

Sterile Procedural Trays, labeled as the following: a. GYN-URO LAPSCP CDS-LF b. ROBOTIC CDS c. GYN LAPAROSCOPY CDS-LF d. GYN LAPAROSCOPY e. GYN LAPAROSCOPY CDS f. UNIVERSAL ROBOTIC ORLANDO g. GYN ROBOTIC CDS h. GYN LAPAROSCOPY CDS-1 i. CUH GYN LAPAROSCOPY CDS j. CUH GEN/GYN ROBOTIC CDS k. LAVH CDS-LF l. GYN LAPAROSCOPY CDS-LF m. GYN LAPAROSCOPY CDS UCC-LF n. DAY SURG GYN LAP CDS o. GYN LAPAROSC

DateFeb 6, 2023 (initiated Nov 17, 2022)
StatusOpen, Classified
ReasonSterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Root causeDevice Design
ActionMedline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Quantity8,064 cases (12,425 units)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.