Recall Z-1077-2023 ยท FTN Plastic Surgery And Accessories Kit

MEDLINE INDUSTRIES, LP - Northfield

Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINOPLASTY CDS d. ACH PLASTICS BREAST-LF e. NE BREAST RECON CDS f. PLASTIC SURGERY CDS g. MINOR PLASTIC CDS h. GROTH BREAST CDS i. BREAST ABD j. MINOR/U-BAR CDS-LF k. PLASTIC BREAST CDS-LF l. BREAST-HERNIA-PORT CDS-LF m. DAY SURG PLASTIC CDS n. PLASTIC SURGERY PACK o. PLASTIC PACK-LF p.

DateFeb 6, 2023 (initiated Nov 17, 2022)
StatusOpen, Classified
ReasonSterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Root causeDevice Design
ActionMedline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Quantity6,420 cases (15,009 units)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.