Recall Z-1074-2023 ยท FDE Laparoscopy Kit

MEDLINE INDUSTRIES, LP - Northfield

Sterile Procedural Trays, labeled as the following: a. LAPAROTOMY CDS-LF b. LAPAROSCOPY CDS-LF c. BASIC LAPAROSCOPY CDS d. LAPAROSCOPY CDS-LF e. GENERAL LAPAROSCOPY PROCEDURE f. LAP CHOLE CDS g. COLO-RECTAL CDS-LF h. MAJOR PROCEDURE-LF i. HYPOSPADIUS CDS j. LAPAROSCOPY-GENERAL k. LAP CHOLE CDS l. GENERAL LAPAROSCOPY CDS m. MINOR PROCEDURE CDS n. MAJOR CDS o. HERNIA CDS p. MINOR CDS q. MAJOR PROCE

DateFeb 6, 2023 (initiated Nov 17, 2022)
StatusOpen, Classified
ReasonSterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Root causeDevice Design
ActionMedline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Quantity42,593 cases (117,470 units)
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.