Recall Z-1053-2023 ยท KPS System, Tomography, Computed, Emission

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING

GE Nuclear Medicine 600/800 series systems, specifically: a) 870, b) 870 DR 3/8 inch NM GANTRY WITH NEW ROTOR - MEA, c) 870 DR / Dismount, d) 870 DR / Dismount / 60HZ, e) 870 DR 3/8 inch WSO, f) 870 DR 3/8 inch NM Gantry, g) 870 DR 3/8 inch, h) 870 DR 3/8 inch CH, i) 870 DR 5/8 inch NM GANTRY, j) 870 DR BJ / Dismount, k) 870 DR CyberDefense 3/8 inch NM Gantry, l) 870 DR NM GANTRY FIN

DateFeb 8, 2023 (initiated Dec 19, 2022)
StatusOpen, Classified
ReasonA mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
Root causeProcess control
ActionGE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION on 12/19/2022 by mail. The notice explained the issue and the risk, and requested the following action by taken: .GE Healthcare asks that you STOP USE of your Nuclear Medicine system, if your system is listed in the Affected Product Details and Appendix of this letter, until an inspection of your system is completed by GE Healthcare and you have been provided with a letter indicating the system can be returned to clinical use. .GE Healthcare will correct all affected products at no cost to you. GE Healthcare will inspect your system t
Quantity557 units
DistributionWorldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Gree
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093514DISCOVERY MODEL NM/CT 670
Ge Medical Systems F.I. Haifa, Dec 10, 2009