Recall Z-1040-2023 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

System Material # Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee multi-purpose 10094139 Artis zee biplane 10094141 Artis zeego 10280959 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zeego 10848283 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Fluoroscopic X-Ray System

DateJan 30, 2023 (initiated Dec 16, 2022)
StatusOpen, Classified
ReasonIn rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
Root causeSoftware design
ActionOn December 16, 2022, the firm initiated the recall to its affected customers via Urgent Medical Device Correction letters. Customers were instructed to do the following: Observe if the system suffers an abrupt shutdown. This can be caused e.g., by usage of the Emergency SHUTDOWN button, manually switching off via End session / Shutdown System command in the main menu, or by a long press on the power-on push button on the system console. In the event of an abrupt shutdown, restart immediately to verify system state and call service if necessary. The firm will provide a software patch to af
Quantity609 US; 1627 worldwide
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018