ClearanceBrief
Recall Z-1024-2023 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, OXY/CVR COMBO BB841 FUSION W/BALANCE 26L, REF BB841. Used in Cardiopulmonary bypass.

DateJan 27, 2023 (initiated Dec 12, 2022)
StatusOpen, Classified
ReasonAn upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Root causeProcess change control
ActionMedtronic initially notified consignees on 12/12/2022 and followed up with an updated notification on 12/21/2022. Consignees were advised of two options: 1. Continue to use the Affinity Fusion Oxygenator without using the TMA (use other conventional perfusion circuit temperature monitoring methods); or 2. Continue to use the Affinity Fusion Oxygenator and the TMA for arterial temperature monitoring. Ensure there is minimal torque applied to the oxygenator TMA when attaching or detaching the Temperature Probe (Product Number ATP210). Also, minimize manipulation of the TMA-probe connection duri
Quantity90,220 units
DistributionWorldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Den
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203111Affinity Fusion Oxygenator System, Affinity Pixie Oxygenation System
Medtronic, Sep 16, 2021