ClearanceBrief
Recall Z-0966-2023 ยท FPA Set, Administration, Intravascular

Smiths Medical ASD Inc.

CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24, 3) REF 21-7302-24, 4) REF 21-7308-24, 5) REF 21-7309-24, 6) REF 21-7310-24, 7) REF 21-7322-24, 8) REF 21-7323-24, 9) REF 21-7324-24, 10) REF 21-7333-24, 11) REF 21-7343-24, 12) REF 21-7346-24, 13) REF 21-7349-24, 14) REF 21-7359-24, 15) REF 21-7363-24, 16) REF 21-7383-24, 17) REF 21-7391-24, 18) REF 21-7

DateJan 30, 2023 (initiated Dec 9, 2022)
StatusOpen, Classified
ReasonThere are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
Root causeUnder Investigation by firm
ActionThe firm issued an URGENT MEDICAL DEVICE CORRECTION notice on 12/09/2022 by mail. Actions for Clinicians and Patients: . The pump may not adequately detect the cassette before or during an infusion due to this issue, and an alarm will be triggered. If a pump displays an NDA alarm, the user can try to resolve the alarm by repositioning the CADD Medication Cassette Reservoir while connected to the pump, repositioning the reservoir by disconnecting from the pump and reattaching it to the pump, or replacing the reservoir. . Or the user can remove the reservoir from the pump and push the plasti
Quantity47595495 units
DistributionWorldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.