Recall Z-1018-2023 ยท HRY Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer

Biomet, Inc.

Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a femoral and tibial component Item Number: 154341

DateJan 26, 2023 (initiated Dec 15, 2022)
StatusOpen, Classified
ReasonPackaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier
Root causePackage design/selection
ActionZimmer Biomet issued Urgent Medical Device Recall Letter on 12/15/22 to Distributors and Risk Managers/Surgeons. Letter states reason for recall, health risk and action to take: Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the notice for hospitals and ens
Quantity7 units
DistributionWorldwide distribution - US Nationwide distribution in the states of IN, KY and the countries of Canada, New Zealand, Japan, Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133940OXFORD FIXED LATERAL BEARING PARTIAL KNEE REPLACEMENT (BEARINGS / TRAYS)
Biomet UK, Ltd., Mar 28, 2014