Recall Z-1180-2023 ยท GAW Suture, Nonabsorbable, Synthetic, Polypropylene

Covidien, LP

Surgipro II Monofilament Polypropylene Sutures Product Description Model Number VP-421X SURGIPRO*II 4-0 90CM V20 DA7P VP421X VP-435-X SURGIPRO II 4-0 90CM KV5 DA7P VP435X VP-541X SURGIPRO*II 4-0 BLU 120CM V20DA VP541X VP-542-X SURGIPRO II 3-0 BLU 120CM V20DA VP542X VP-735-X SURGIPRO*II 7-0 60CM MV1758DA VP735X VP735X-12 SURGIPRO*II 7-0 60CM MV1758 DA VP735X12 VP-738-X SURGIPRO*II 8-0BLU 60CM CV351

DateFeb 25, 2023 (initiated Dec 2, 2022)
StatusOpen, Classified
ReasonSterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
Root causePackaging process control
ActionMedtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected
Quantity322152 units
DistributionUS Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Geor
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050947MODIFICATION TO: MODIFIED USS POLYPROPYLENE SUTURE