ClearanceBrief
Recall Z-0959-2023 ยท BSZ Gas-Machine, Anesthesia

Getinge Usa Sales Inc

Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400

DateJan 26, 2023 (initiated Dec 22, 2022)
StatusOpen, Classified
ReasonDue to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
Root causeSoftware Design Change
ActionOn December 22, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to examine their inventory immediately to determine if they have any of the affected serial numbers. If you have an affected system running on software version 4.8.0 through 4.8.3, you may continue use of the device until the software is updated with the following considerations: 1. If during use, the device is switched from Automatic to Manual and it is found that the manual bag is empty the user can activate the Emergency ventilation or turn off/on the devi
Quantity8,016 Total (7,766 OUS; 250 US)
DistributionUS nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System
Maquet Critical Care AB, Mar 18, 2020