Recall Z-1007-2023 ยท HQC Unit, Phacofragmentation

Johnson & Johnson Surgical Vision, Inc.

Single-Use Packs, REF: OPO80

DateJan 20, 2023 (initiated Dec 7, 2022)
StatusOpen, Classified
ReasonWeld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Root causeProcess control
ActionOn 12/07/22, recall notices were sent to customers who were asked to do the following: 1) Discontinue using affected product. 2) Return product to the recalling firm. 3) Complete and return the Customer Reply form via email to RegCompliOne@its.jnj.com 4) Distributors, distribute the recall notice to your customers. Customers with questions or concerns can call the firm at 1-877-266-4543 Option 2.
Quantity384
DistributionUS: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133115COMPACT INTUITIV; SYSTEM, WIRELESS REMOTE CONTROL, FOUR-BUTTON FOOT PEDAL, SINGLE-USE FLUIDICS PACK
Abbott Medical Optics, Inc., Feb 27, 2014