Recall Z-1172-2023 ยท GAR Suture, Nonabsorbable, Synthetic, Polyamide

Covidien, LP

Dermalon Monofilament Nylon Suture: (1) 88861741-11 DERMALON* 6-0 BLU 45CM C1 (2) 88861747-11 DERMALON* 6-0 BLU 45CM C12 (3) 88861749-24 DERMALON* 5-0 BLU 45CM P13 (4) 88861756-11 DERMALON* 6-0 BLU 45CM C13 (5) 88861757-41 DERMALON* 3-0 BLU 75CM C13

DateFeb 25, 2023 (initiated Dec 2, 2022)
StatusOpen, Classified
ReasonSterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
Root causePackaging process control
ActionMedtronic issued an "URGENT: MEDICAL DEVICE RECALL" letter to US consignees on 01-Dec-2022 via UPS 2-day delivery. OUS countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and actions to take: Actions: " Identify and quarantine all unused and non-expired affected non-absorbable sutures. " Please note the affected product may be located within Custom Suture Packs. Please reference Custom Suture Packs to identify affected product. " Please complete the enclosed Customer Confirmation Form (even if your account no longer has affected
Quantity9606
DistributionUS Nationwide Distribution and OUS Foreign distribution to countries of: Albania Argentina Armenia Australia Austria Azerbaijan Belarus Belgium Bosnia And Herzegovina Brazil Bulgaria Cambodia Cameroon Canada Canary Islands Chile China Colombia Croatia Czech Republic Denmark Egypt Finland France Geor
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981582SURGILON, DERMALON, & OPHTHALON
Davis & Geck, Inc., Jul 13, 1998