Recall Z-1017-2023 ยท PNN Orthodontic Software

Align Technology Inc

Invisalign Express 10 REF 8512 Invisalign System Comprehensive REF 9000 Invisalign System Moderate REF 9001

DateJan 26, 2023 (initiated Dec 5, 2022)
StatusCompleted
Reason3D orthodontic planning software has a defect that leads to an issue where incorrect number of aligners may be produced (less or greater than the doctor requested) and incorrect packaging is provided.
Root causeSoftware design
ActionOn 12/05/2022, the firm initiated telephone calls with customer to inform them of a voluntary recall of selected Invisalign aligners. Starting on 12/09/2022, the firm sent an "URGENT Medical Device Correction" email to customers as a follow-up to a telephone call informing them that selected Invisalign aligners may have the potential clinical impact and therefore may result in unnecessary/unprescribed treatment due to software issue with the new Treatment Planning History feature within ClinCheck software. Customers are instructed: 1. Cease/stop using and dispose of the affected Invisalign ca
Quantity14 cases (637 aligner and temples)
DistributionU.S. Nationwide distribution in the states of AZ, CA, CO, FL, KY, MA, NJ, NY, OH, OK, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220287Invisalign System
Align Technology, Inc., Apr 7, 2022