Recall Z-1038-2023 ยท FRG Wrap, Sterilization

Turbett Surgical, Inc.

Turbett Surgical Container, TS1000: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

DateJan 30, 2023 (initiated Dec 5, 2022)
StatusOpen, Classified
ReasonPod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.
Root causeDevice Design
Action***UPDATE*** Turbett Surgical issued via E-mails Product Alert directly to the Hospital departments utilizing the device on 12-27-2022. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/User: 1. Immediately discontinue the use of the specific devices exhibiting one or more failure(s) listed below. " Missing side supports " Broken or cracked side tab support " Backwall weld separation " Side weld separation " Crack in the backwall seam 2. Email (quality@turbettsurgical.com) with the following information. " Hospital name " Serial number
Quantity6 units
DistributionUS Nationwide distribution in the states of CA, IA, IN, NC, NH, NY, OH, LA, SD, TX, VA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153025Turbett Surgical Container
Turbett Surgical, LLC, Jan 14, 2016