ClearanceBrief
Recall Z-0949-2023 · MSW System, Test, Thyroglobulin

Siemens Healthcare Diagnostics, Inc.

IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/ IMMULITE 2000 XPi) for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # L2KTY2(D) Siemens Material Number (SMN): 10381647

DateJan 11, 2023 (initiated Dec 7, 2022)
StatusOpen, Classified
ReasonPotential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe increased imprecision with low level patient samples
Root causeUnder Investigation by firm
ActionSiemens Healthcare issued An Urgent Medical Device Correction (UMDC) to US affected customers by FedEx beginning 12/08/2022. An Urgent Field Safety Notice (UFSN) was issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 12/07/2022. Letter states reason for recall, health risk and action to take: Discontinue use and discard Thyroglobulin kit lots listed in Table 1. Please review this letter with your Medical Director, including consideration for review of previously generated results as outlined
Quantity717 kits
DistributionNationwide Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Bangladesh¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bolivia¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil¿¿¿¿¿¿¿¿¿ Bulgaria¿¿¿¿¿¿¿ Canada¿¿¿¿¿¿¿¿¿ Chile¿¿¿¿¿¿¿¿¿¿ Colombia¿¿¿¿¿¿¿ Costa¿Rica¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Dominica
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982468IMMULITE 2000 THYROGLOBULIN MODEL L2KTY
Diagnostic Products Corp., Mar 12, 1999