ClearanceBrief
Recall Z-0914-2023 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Vanta/Seqenita LT Clinician Programmer Application, Neurostimulation systems for pain therapy, Model Number CT900D

DateJan 6, 2023 (initiated Dec 3, 2022)
StatusOpen, Classified
ReasonSoftware anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
Root causeUnder Investigation by firm
ActionThe firm issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/03/2022 by mail. The notices explained the problems and the associated risks and provided the following: "Recommended Mitigation If the INS cannot be interrogated with Vanta PP App, contact Medtronic Technical Services at 1-800-707-0933. If needed, Medtronic will schedule an in-field service appointment with the managing SCS physician and the patient to diagnose and reset the INS." "Medtronic is working on a Vanta CP App software update that will resolve this issue and will communicate additional informat
Quantity1337 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.