Recall Z-0913-2023 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Medtronic Neuromodulation
Vanta/Seqenita LT Clinician Programmer Application, Neurostimulation systems for pain therapy, Model Number CT900C
| Date | Jan 6, 2023 (initiated Dec 3, 2022) |
| Status | Open, Classified |
| Reason | Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303" |
| Root cause | Under Investigation by firm |
| Action | The firm issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/03/2022 by mail. The notices explained the problems and the associated risks and provided the following: "Recommended Mitigation If the INS cannot be interrogated with Vanta PP App, contact Medtronic Technical Services at 1-800-707-0933. If needed, Medtronic will schedule an in-field service appointment with the managing SCS physician and the patient to diagnose and reset the INS." "Medtronic is working on a Vanta CP App software update that will resolve this issue and will communicate additional informat |
| Quantity | 18 units |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.