Recall Z-1103-2023 ยท JAK System, X-Ray, Tomography, Computed

Olympus Corporation of the Americas

SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)

DateFeb 7, 2023 (initiated Dec 13, 2022)
StatusOpen, Classified
ReasonA grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged procedure/surgery due to additional time to troubleshoot the device
Root causeUnder Investigation by firm
ActionOlympus issued URGENT - Medical Device Field Corrective Action letter on 12/13/22. Letter states reason for recall, health risk and action to take: 1. Notify your users about this issue and stop using the single-use bronchoscopes (models INS-7100 or INS-7130) with SYS-4000 system until cable is replaced. Your users can continue to use the SYS-4000 using a reusable bronchoscope, with its respective video processor, without affecting their electromagnetic navigation (EMN) procedural experience. 2. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal
Quantity13 units
DistributionFL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170023SPiN Thoracic Navigation System
Veran Medical Technologies, Inc., May 15, 2017