Recall Z-1009-2023 ยท OOI Real Time Nucleic Acid Amplification System

Abbott Molecular, Inc.

Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.

DateJan 20, 2023 (initiated Dec 9, 2022)
StatusOpen, Classified
ReasonThe firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect
Root causeUnder Investigation by firm
ActionOn December 09, 2022, Abbott Molecular issued a Field Correction Recall/Urgent Field Safety Notice to all customers who have the Alinity m System by FedEx priority overnight express. The letter explained the problems and provided the following information: "An assessment of your Alinity m System associated with the above listed potential issues will be completed remotely if your system is currently connected to AbbottLink or alternatively via onsite inspection by an Abbott representative. If necessary, an Abbott representative will schedule time to corrected the affected ADU9s) on your Ali
Quantity918 units
Distributionx
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.