Recall Z-0775-2023 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

Programmable Diagnostic Computer

DateJan 1, 2023 (initiated Nov 16, 2022)
StatusOpen, Classified
ReasonThe firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during system boot; 2) Subsystem crash during examination; 3) Dialog Monitor Computer (DMC) application crash while loading a study; and 4) Software crash due to system internal timeout. Issue 1 may lead to a delay or interruption of procedure. Issues 2, 3, and 4 may result in delay in starting or continuing the examination, and may also prevent the operator from starting or con
Root causeSoftware design
ActionOn November 16, 2022, the firm notified customers with an Urgent Medical Device Correction letter delivered electronically via EchoSign. Customers were informed of the 4 software issues. Siemens will update the software in affected systems to resolve the issues. The firm's service organization will contact customers to arrange a date to perform the corrective action. You may contact the firm's service organization for an earlier appointment at 1--800-888-7436.
Quantity2213 distributed worldwide; 638 US
DistributionDomestic distribution nationwide. Worldwide foreign distribution to Albania Algeria Angola Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belgium Bolivia Bosnia and Herzegovina Botswana Bulgaria Cambodia Canada Chile China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark
FDA recordOpen on fda.gov

510(k) clearances named in this recall