Recall Z-0764-2023 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

Revaclear 400 dialyzers. Indicated for treatment of chronic and acute renal failure by hemodialysis.

DateDec 22, 2022 (initiated Nov 23, 2022)
StatusOpen, Classified
ReasonReports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak
Root causeProcess control
ActionBaxter Healthcare notified healthcare facilities on 11/23/2022 and home patients on 11/29/2022, via first class mail. Consignees were instructed to be aware of the difference in appearances between Polyurethane Displacement and an Internal Blood Leak and ensure staff is also aware. Additionally, consignees were instructed that if there is a blood leak alarm, to follow your clinical practice to stop treatment, and if only the Polyurethane Displacement is observed, to continue use of the dialyzer. Consignees were asked to acknowledge receipt of the notification by completing and returning the
Quantity1,140,096 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130039REVACLEAR 300 DIALYZER, REVACLEAR 400 DIALYZER
Gambro Renal Products, Inc., May 2, 2013