Recall Z-0760-2023 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

Revaclear 400 Dialyzer. Product Code: 114746L. Used to treat chronic and acute renal failure by hemodialysis.

DateDec 22, 2022 (initiated Nov 23, 2022)
StatusOpen, Classified
ReasonPotential for Internal Blood Leaks due to a twisted gasket in the dialyzer
Root causeProcess change control
ActionBaxter Healthcare sent a notification via USPS first class mail on about 11/23/2022 and to home patients on about 11/29/2022. Consignees were instructed to locate and remove all affected product from your inventory, contact Baxter Healthcare Center for Service to arrange for return and credit, and acknowledge the receipt of the notification by completing a reply form on the customer portal. Note: this letter describes two issues. This recall addresses Internal Blood Leaks being caused by a twisted gasket in the dialyzer. The second issue is linked to reports of Polyurethane Displacement in
Quantity3,610,944 units
DistributionWorldwide distribution. US Nationwide including Guam and St. Thomas USVI; Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecua
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130039REVACLEAR 300 DIALYZER, REVACLEAR 400 DIALYZER
Gambro Renal Products, Inc., May 2, 2013