ClearanceBrief
Recall Z-0908-2023 · HGP Electrode, Circular (Spiral), Scalp And Applicator

Philips North America Llc

Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631

DateJan 6, 2023 (initiated Nov 21, 2022)
StatusOpen, Classified
ReasonFetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of this product.
Root causeUnder Investigation by firm
ActionPhilips Medical issued to U.S. customers, the Urgent Medical Device Correction (UMDC) letter via certified mail 11/21/22. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take. Philips has decided to discontinue the distribution of this product. To limit the potential for the spiral tip of the FSE to break off in the patient s scalp, it is important to avoid over rotation during attachment or pulling the tip from the fetal scalp Customer should complete the Urgent Field Safety Notice Respo
Quantity69,679 cases (1,741,975 units)
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Croatia, Cura¿ao, Czechia, Denmark, Faroe Islands, Finland, France, Germany, Hong Kong, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Luxembourg, Malaysi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030691FETAL SPIRAL ELECTRODE, MODEL FSE07000
Clinical Innovations, Inc., Sep 17, 2003