Recall Z-0877-2023 ยท NIE Catheter, Hemodialysis, Triple Lumen, Non-Implanted

Covidien LP

MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No. 8888135131; 13.5 cm, Curved Extensions, Kit - Model No. 8888135132; 13.5cm, Pre-Curved, Kit - Model No. 8888135133; 13.5cm, Straight Extensions, IC Tray, Model No. 8888135134; 16cm, Straight Extensions, Kit - Model No. 8888135161; 16cm, Curved Extensions,

DateJan 12, 2023 (initiated Dec 7, 2022)
StatusOpen, Classified
ReasonThere is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Root causeProcess design
ActionOn December 7, 2022, Medtronic began notifying customers who had received the affected products. Customers were instructed to quarantine, discontinue use, and return unused products. Customers should forward the notification to those who need to be aware within the organization, and to any location where the devices have been distributed.
Quantity359,730 (23,003 US; 336,727 OUS)
DistributionWorldwide distribution. US nationwide including Guam, USVI, and Puerto Rico; Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Polynesia, Germany, Greece, Hungary, Iceland, Italy, Jamaica, K
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030209MAHURKAR OPLUS CATHETER, MODEL 13.5 FR
The Kendall Company, May 13, 2003