Recall Z-0907-2023 ยท PLS Immunohistochemistry Assay, Antibody, Programmed Death-Ligand 1
Ventana Medical Systems, Inc.
VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001
| Date | Jan 6, 2023 (initiated Nov 18, 2022) |
| Status | Open, Classified |
| Reason | Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result. |
| Root cause | Process design |
| Action | On 11/18/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION", Letter via UPS ground to customer to inform them that internal comparison studies revealed unacceptable, light staining with some VENTANA PD-L1 (SP142) lots. This can cause a reduction in both staining intensity and percent positivity, which could potentially lead to a false-negative PD-L1 result. Actions required of Customers: 1) Immediately discontinue the use of the VENTANA PD-L1 (SP142) lot numbers listed in Table 2 and discard any remaining inventory according to your local guidelines. 2) Be aware of the new expiration |
| Quantity | 8,927 kits |
| Distribution | Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Cze |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.