Recall Z-0460-2023 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

Philips Ingenia Elition X MR System, Model Numbers 781358, 782107, 782119, 782136

DateDec 8, 2022 (initiated Nov 22, 2022)
StatusOpen, Classified
ReasonGradient Coil may act as a heat source with a potential to produce smoke and/or fire.
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Correction Notice dated 11/22/22 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users 1. Systems that do not yet have the SmokeDetector Interlock feature installed are identified in Appendix A. If your system is listed in Appendix A immediately discontinue use of your impacted MR system(s) until Philips has installed the SmokeDetector Interlock. " Post a Do not use notice on or near the impacted MR system(s) " Post this notice near the affected MR system(s) for ease of reference. " Circulate this
Quantity566 total
DistributionUS Nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173356Ingenia Elition S and Ingenia Elition X R5.4
Philips Medical Systems Nederland B.V., Mar 6, 2018
K173451Ingenia Elition S and Ingenia Elition X
Philips Medical Systems Nederland B.V., Mar 20, 2018
K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., Feb 14, 2019
K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., Apr 10, 2020
K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi Nudclei
Philips Medical Systems Nederland B.V., Mar 3, 2022