ClearanceBrief
Recall Z-0738-2023 ยท MQP Spinal Vertebral Body Replacement Device

NuVasive Inc

X-CORE 2 Ti Core, Static. Part of the NuVasive X-Core Expandable VBR System used with spinal fixation.

DateDec 20, 2022 (initiated Nov 10, 2022)
StatusOpen, Classified
ReasonPotential that insert will be unable to disengage from the vertebral body replacement device/implant
Root causeDevice Design
ActionOn 11/10/2022, the firm sent an "URGENT FIELD SAFETY NOTICE" Letter via email to customers informing them that the firm has received product complaints regarding difficulty in disengaging the Inserter from the X-Core, Static Core implant. NuVasive is updating the Instructions For Use (IFU) and Surgical Technique Guide (STG) to clarify the best fit inserter with each core which can help prevent the user experiencing interference when removing the inserter from the Static Core. Recommended User Actions: This FSN details updates to the IFU document that physicians should consult prior to and
Quantity2,031 devices
DistributionWorldwide - U.S. Nationwide distribution including in the states of NC, CA, FL, KY, IL, TX, CO, OH, AK, GA, UT, MA, WA, HI, AL, NY, WI, AR, ID, VA, PA, SC, MD, PA, AZ, RI, IN, ME, NJ, MI, TN, MN, OR, OK, WY, and Puerto Rico. The countries of Australia, Belgium, Brazil, Colombia, Germany, Italy, Spai
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090176NUVASIVE ADJUSTABLE VERTEBRAL BODY REPLACEMENT SYSTEM
Nuvasive, Inc., Apr 27, 2009