Recall Z-0480-2023 ยท NEW Suture, Surgical, Absorbable, Polydioxanone

Ethicon, Inc.

PDS II (polydioxanone) Suture- VIOLET MONOFILAMENT 27IN 0 S/A CT-1 Product Code: Z340H For use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery

DateDec 14, 2022 (initiated Nov 16, 2022)
StatusOpen, Classified
Reasonknot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage
Root causeUnder Investigation by firm
ActionEthicon notified the four impacted countries via "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" beginning Nov 16, 2022. Letter states reason for recall, health risk and actions to be taken. The customers were instructed to: Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded
Quantity7,560 eaches
DistributionInternational (OUS) only: Canada, Japan, Australia and New Zealand
FDA recordOpen on fda.gov