Recall Z-0885-2023 · DSP System, Balloon, Intra-Aortic And Control
Datascope Corp.
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-32, 0998-UC-0800-33, 0998-UC-0800-34, 0998-UC-0800-35, 0998-UC-0800-36, 0998-UC-0800-45, 0998-UC-0800-52, 0998-UC-080
| Date | Jan 19, 2023 (initiated Dec 19, 2022) |
| Status | Open, Classified |
| Reason | The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter. |
| Root cause | Device Design |
| Action | An URGENT MEDICAL DEVICE CORRECTION notification letter dated 12/19/2022 was sent to customers. It contains clinical guidance to augment current clinical recommendations of patient management strategies should an IAB perforation occur. Maintain established patient and device management strategies regarding timing to IAB catheter removal. User Actions to be taken include to examine inventory immediately to determine if affected units are on hand, if a blood back event is suspected, remove the Cardiosave from patient use, and report the event to appropriate biomedical engineering staff for insp |
| Quantity | 11,906 Total |
| Distribution | Worldwide Distribution - US Nationwide and the countries of Albania, Angola, Argentina, Australia, Austria, AZERBAIJAN, BAHRAIN, BANGLADESH, Barbados, Belarus, Belgium, BOSNIA and HERZEGOVINA, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d'Ivoire, Croatia, CZECH REPUBLI |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181122 | CARDIOSAVE Intra-Aortic Balloon Pump Datascope Corp., May 31, 2018 |