Recall Z-0344-2023 ยท JAK System, X-Ray, Tomography, Computed

Mobius Imaging, LLC

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

DateDec 6, 2022 (initiated Nov 18, 2022)
StatusTerminated
ReasonStryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.
Root causeOther
ActionOn November 18, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed that a service representative will contact their facility to schedule a time to replace the batteries in the affected devices. The customer does not need to take additional action while waiting for the battery preplacement. If you have any questions or concerns regarding this update, please contact shirleytechnicalsupport@stryker.com or call Stryker Technical Support at 978 386 9619.
Quantity3
DistributionDomestic distribution to consignees in Colorado, Tennessee, and Utah.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180393AIRO Computed Tomography (CT) X-ray System
Mobius Imaging, LLC, Mar 15, 2018