ClearanceBrief
Recall Z-0730-2023 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Medical Systems, LLC

B125P Patient Monitor (with E-module slot option), REF 6160000-002

DateDec 19, 2022 (initiated Nov 16, 2022)
StatusOpen, Classified
ReasonGE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.
Root causeDevice Design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/16/2022. The notice explained the problem and safety issue and requested the following: "You can determine whether your monitor is impacted by looking at the side of your monitor where the parameter acquisition connectors are located. B1x5P with E-module slot option and all B1x5M patient monitors are impacted (See Figure 1). B1x5P patient monitors without E-module slot option are not impacted (See Figure 2). You can also determine whether your monitor is impacted following the affected product direction
Quantity388 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213490Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor B125P
Ge Medical Systems Information Technologies, Inc., Apr 1, 2022