Recall Z-0728-2023 · NKB Thoracolumbosacral Pedicle Screw System

Icotec Ag

VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

DateDec 19, 2022 (initiated Nov 3, 2022)
StatusTerminated
ReasonDuring internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
Root causeNonconforming Material/Component
ActionIcotec issued a Field Action Notice to its consignee on 11/03/2022 by email. The notice explained the issue and that XXXX. The firm followed with an email to inform the consignee of their intent to recall the device. The notice requested the return of the device from the field. The firm will be replacing the devices.
Quantity6 devices
DistributionUS Nationwide distribution in the state of GA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190545VADER® one Pedicle System MIS and LightMore® Pedicle System 6.0
Icotec AG, Jun 20, 2019
K193423VADER®one Pedicle System MIS and LightMore® Pedicle System 6.0
Icotec AG, May 22, 2020
K200596G21 Cement, VADER® Pedicle System
Icotec AG, Oct 13, 2020