ClearanceBrief
Recall Z-0849-2023 ยท JWY Manual Antimicrobial Susceptibility Test Systems

Biomerieux Inc

ETEST CLINICAL IMIPENEM RELEBACTAM IPR US S30, CATALOG 420927

DateJan 4, 2023 (initiated Nov 9, 2022)
StatusOpen, Classified
ReasonAs the temperature and time out of range was exceeded, product performances cannot be guaranteed
Root causeStorage
ActionBiomerieux sent a FIELD SAFETY CORRECTIVE ACTION notice to the US subsidiary on 11/09/2022 by email. The notice requested the subsidiary identify their customers and issue an URGENT FIELD SAFETY NOTICE which explains the problem and request the following actions: "1. Confirm this letter has been distributed to and reviewed by all appropriate personnel within your organization. 2. Check your inventory for any of the impacted products (reference Table 1) associated with local shipments. 3. Stop using and scrap any remaining impacted products in your inventory. a. Request product replacement o
Quantity2 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191953ETEST Imipenem/Relebactam (IPR) (0.002/4-32/4 ug/mL)
Biomerieux S.A., Aug 22, 2019