Recall Z-0479-2023 ยท FPO Stretcher, Wheeled

Stryker Medical Division of Stryker Corporation

Power-PRO 2, Model No. 6507, Item No. 650705550001 (US); 650700000000 (International)

DateDec 14, 2022 (initiated Nov 11, 2022)
StatusOpen, Classified
ReasonCot may experience unintended motion due to damaged cables. Potential risk of soft tissue injury to patients such as scrapes or bruises resulting from the cot lowering faster than expected.
Root causeUnder Investigation by firm
ActionOn November 11, 2022, the firm notified affected customers via "URGENT: MEDICAL DEVICE CORRECTION" letters. Devices require servicing. Customers were instructed to locate affected units and identify the address where they can be serviced. Until servicing can be performed, customers were provided with additional instructions and precautions to follow to ensure safe continued use of the device. Return the enclosed business reply form to confirm receipt of this notification by fax +1 269 488 8691 or email productfieldaction@stryker.com. Upon receipt of the completed business reply form, Stryker
Quantity465 (403 US; 62 OUS)
DistributionWorldwide Distribution: US (Nationwide) and Foreign to countries of: Australia, Canada, Netherlands, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.