ClearanceBrief
Recall Z-0721-2023 ยท LLZ System, Image Processing, Radiological

GE Medical Systems, LLC

Centricity PACS-IW with Universal Viewer version 5.0. Used to display medical images (Including mammograms) and data from various imaging sources.

DateDec 16, 2022 (initiated Oct 28, 2022)
StatusOpen, Classified
ReasonThere is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.
Root causeSoftware Manufacturing/Software Deployment
ActionGE Healthcare notified consignees on about 10/28/2022 via letter titled, "URGENT MEDICAL DEVICE CORRECTION - UPDATE." Consignees were instructed that the device can continued being used by following these instructions: 1) When a study is not present on the Study List, contact a GE Healthcare Service Representative for assistance in recovering the study. 2) In situations where study interpretation is time sensitive, interpret the study at the scanning unit or plan for an alternate means of reading studies for emergent cases. GE Healthcare will contact consignees to arrange for the correction.
Quantity1 unit
DistributionNationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND, NJ, NY, OK, OR, SD, TX, WI, WV, WY. International distribution to Austria, Brazil, France, Germany, Ghana, Ireland, Italy, Kuwait, Poland, Saudi Arabia, South Africa, Spain, Switzerland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123174CENTRICITY PACS-IW WITH UNIVERSAL VIEWER
GE Healthcare, Nov 16, 2012