Recall Z-0741-2023 ยท NDC Calculator, Drug Dose
Companion Medical, Inc.
InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA.
| Date | Dec 20, 2022 (initiated Nov 8, 2022) |
| Status | Open, Classified |
| Reason | Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia. |
| Root cause | Software design |
| Action | On 11/8/22, Medtronic emailed correction notices to customers who were asked to do the following: 1) Verify the running app by: a) Open the app on the device, b) From Home screen tap the "Settings" button, c) Select the "Help and Support" option, d) Under Contact Us section, note the app version listed. 2) If affected app versions are not listed, no action is necessary. 3) If affected apps are listed, uninstall the app and reinstall the version currently available in the app store (5.5.4.4) by following these steps: a) Perform an app search for InPen, b) Tap the icon to the left of the Open bu |
| Quantity | 140 |
| Distribution | US Nationwide distribution including in the states of CA, NY, MT, OH, IN, MN, PA, CT, TX, WA, TN, SD, MI, FL, OR, NC, KY, WI, MO, WV, NV, SC, AZ, AR, MA, LA, VT, GA. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K201337 | InPen System Companion Medical, Inc., Jun 18, 2020 |