Recall Z-0734-2023 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Holdings, Inc.

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

DateDec 20, 2022 (initiated Jul 27, 2022)
StatusOpen, Classified
Reason(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".
Root causeProcess design
ActionFresenius issued Updated Urgent Medical Device Correction REV B Customer Notification letter (Field Action Number: FA-2020-09-W _REV B) to consignees that have CDX equipped machines on 07-27-2022 via USPS First Class Mail. Letter states reason for recall, health risk and action to take: First, a software issue in SW versions 2.72 and earlier. All machines should be updated to current software (SW version 2.74), pursuant to the July 20, 2021, letter. Second, hardware issue that affects only CDX machines. To reduce the chance of EMI causing a false Remove USB Device 2 alarm from occurring we su
Quantity2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205
DistributionUS Nationwide distribution.