Recall Z-0757-2023 ยท ITX Transducer, Ultrasonic, Diagnostic

GE Medical Systems, LLC

Vivid q ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, an

DateDec 22, 2022 (initiated Nov 10, 2022)
StatusOpen, Classified
ReasonPotential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.
Root causeDevice Design
ActionGE Healthcare notified consignees on about 11/10/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION. For U.S. customers, letters were sent via FedEx. Consignees were instructed to follow updated safety instructions provided in the appendix included with the notification letter, place the appendix with the product labeling, and replace the batter every 2 years or if the battery is not capable of powering the system for more than 30 minutes. In response, GE Healthcare requested consignees return the response form.
Quantity2,609 units
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082374VIVID I AND VIVID Q
Ge Medical Systems Israel, Ultrasound, Ltd., Sep 2, 2008
K092140GE VIVID I DIAGNOSTIC ULTRASOUND, GE VIVID Q DIAGNOSTIC ULTRASOUND
Ge Medical Systems Israel, Ultrasound, Ltd., Aug 21, 2009
K102388VIVID I DIAGNOSTIC ULTRASOUND SYSTEM
Ge Medical System Israel , Ltd., Nov 4, 2010
K121062VIVID I DIAGNOSTIC ULTRASOUND SYSTEM, VIVID Q DIAGNOSTIC ULTRASOUND SYSTEM
Ge Medical System Israel , Ltd., Aug 17, 2012