ClearanceBrief
Recall Z-0497-2023 ยท PWL Oral Surgery Tray

Stradis Medical, LLC dba Stradis Healthcare

Surgical room, DENTAL SURGICAL PACK, Item No.41236SDSP

DateDec 14, 2022 (initiated Nov 2, 2022)
StatusOpen, Classified
ReasonMedical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely sealed, which may result in a breach in the sterility of the kit.
Root causePackaging process control
ActionStradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your cus
Quantity20 units
DistributionUS nationwide distribution, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.