ClearanceBrief
Recall Z-0420-2023 · NKB Thoracolumbosacral Pedicle Screw System

Integrity Implants Inc.

LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Labeled as the following sizes: a. 5.5mm x 35mm; b. 5.5mm x 40mm; c. 5.5mm x 45mm; d.

DateDec 5, 2022 (initiated Jan 27, 2022)
StatusTerminated
ReasonPotential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.
Root causeDevice Design
ActionIntegrity Implants notified consignees on 01/28/2022, via emailed letter. The letter instructed consignees to quarantine affected product, complete and return the Product Tracking & Verification Form, and return affected product. On 02/05/2022 the recall was expanded, and consignees were notified via emailed letter. The listed the additional affected product and contained the same instructions to consignees.
Quantity1,369 screws
DistributionNationwide distribution to AZ, CA, CT, DC, FL, IL, IN, LA, MA, MN, NJ, OK, PR, TN, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203367LineSider™ Spinal System
Integrity Implants, Inc., Dec 3, 2020