Recall Z-0325-2023 ยท OGJ Nerve Block Tray
Busse Hospital Disposables, Inc.
Busse JOINT INJECTION TRAY -Intended as a nerve block REF No. 8561 R1
| Date | Nov 29, 2022 (initiated Oct 21, 2022) |
| Status | Open, Classified |
| Reason | Kits containing 3M Steri-Drape Surgical Drapes recalled by 3M due to liner on the adhesive component is difficult to remove without damaging the product and may render the product unusable |
| Root cause | Device Design |
| Action | Busse issued Product Recall Notification Letter dated 10/20/22 to Distributor; End User via Email on 10/21/2022 and Certified Mail on 10/24/2022 Letter states reason for recall, health risk and action to take: 1. Immediately check your inventory for the lots listed above. 2. Do not use the drape included in the product with the lot number(s) listed above. 3. Complete the enclosed Response Form within the next 10 days and return via email to csabillon@busseinc.com even if you have no affected product on hand. a. If you do have product in stock, destroy the drape following your facility policy. |
| Quantity | 80 cases |
| Distribution | TX |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.