ClearanceBrief
Recall Z-0328-2023 ยท JWY Manual Antimicrobial Susceptibility Test Systems

Remel Inc

Thermo SCIENTIFIC, Sensititre GN6F, Gram Negative IVD AST

DateDec 8, 2022 (initiated Oct 20, 2022)
StatusTerminated
ReasonAn internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.
Root causeProcess control
ActionThe firm issued an "Urgent Medical Device Correction" notice to its consignees on 10/25/2022 by US mail. The notice explained the issue, risk to health, and notified the consignees of an update to the IFU, specifically, "to add limitations for Carbapenems (Imipenem, Doripenem, Meropenem, Ertapenem), Cefepime, Piperacillin/Tazobactam and Aztreonam added for users to "NOT REPORT" any result for each member of the Protea tribe (including Proteus spp., Providencia spp., Morganella Morganii) to address false susceptible results." Please complete and return the Required Customer Reply form via ema
Quantity2471 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060621SUSCEPTIBILITY TEST PLATE FOR ERTAPENEM 0.002-16UG/ML FOR GRAM NEGATIVE. SUSCEPTIBILITY TEST PLATE FOR ERTAPENEM 0.015-
Tds Holdings, Inc., Apr 21, 2006
K073424SENSITITRE 18-24 HOUR SUSCEPTIBILITY PLATES
Trek Diagnostic Systems, Inc., Jan 22, 2008
K860753SENSITITRE SUSCEPTIBILITY PLATES
Gibco Laboratories Life Technologies, Inc., May 6, 1986
K880641SENSITITRE STABILITY PLATES
Sensititre , Ltd., May 13, 1988
K936305SENSITITRE/PIP/TA2 SUSCEPTIBILITY PLATES
Radiometer America, Inc., Jul 13, 1995
K983244SENSITITRE 18-24 HR SUSCEPTIBILITY PLATES(ADDITION OF MEROPENEM)
Trek Diagnostic Systems, Inc., Nov 30, 1998