Recall Z-0352-2023 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

ARROW INTERNATIONAL Inc.

Pressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-PCAM1, g) ASK-45703-PCCH2, h) ASK-45703-PN; catheter introducer

DateDec 10, 2022 (initiated Oct 27, 2022)
StatusTerminated
ReasonThere is the potential for inadequate connection between the top and bottom housings of the Micro Clave Clear Connectors included in the Arrow MAC Two-Lumen Central Venous Access Kit and the Arrow Pressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit.
Root causeProcess design
ActionThe firm issued an "URGENT MEDICAL DEVICE RECALL" to its consignees on 10/27/2022 by FedEx 2-day mail. The notice explained the problem, the risk, and requested the following: "Action list number 1 Medical facilities 1. Immediately check your inventory for product within the scope of this recall. Users should cease use and distribution of affected product and immediately quarantine the affected product. 2. If you have affected product, complete the RECALL ACKNOWLEDGEMENT FORM, Contact Teleflex Customer Service to receive a return goods authorization (RGA) number. 3. Complete and return form
Quantity1275 units
DistributionUS Distributor in states of: FL, GA, KY, OH, PA and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002507TWO-LUMEN CENTRAL VENOUS ACCESS KIT WITH HEMOSTASIS VALVE/SIDE PORT
Arrow Intl., Inc., Mar 21, 2001
K071538ARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER
Arrow International, Inc., Aug 30, 2007