Recall Z-0340-2023 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

INTUITIVE Da Vinci Xi 8mm SureForm 30 Curved-Tip Stapler Da Vinci Xi|X REF 488530 INTUITIVE Da Vinci 8mm SureForm 30 Reload 30 mm Gray 4-Row 2.0 mm Xi|X REF 48230M INTUITIVE Da Vinci 8mm SureForm 30 Reload 30 mm White 4-Row 2.5 mm Xi|X REF 48230W INTUITIVE Da Vinci 8mm SureForm 30 Reload 30 mm Blue 4-Row 3.5 mm Xi|X REF 48230B

DateNov 30, 2022 (initiated Oct 7, 2022)
StatusOpen, Classified
ReasonPotential staple deployment failure and device fragment generation
Root causeNonconforming Material/Component
ActionOn 10/07/2022, Intuitive Surgical communicated an "Urgent Product Safety Notice" email to "STOP USE of da Vinci Xi 8mm SureForm 30 Stapler Instruments (488530-10) and Reloads (48230M-02, 48230W-02, 48230B-02)" and informing customers that there is a potential to transect targeted tissue without stapler formation that could result in bleeding and potential to generate fragments when using these products. On or about 10/25/2022, the firm sent an "Urgent Medical Device Removal" Letter via FedEx that informed their customers of the STOP USE. Customers are instructed to: Locate and return all aff
Quantity160 boxes of 488530-10 = 960 individual instruments 602 reload boxes = 7224 individual reloads
DistributionU.S.: Arizona, California, Colorado, Florida, Illinois, Indiana, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Nevada, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, and Wisconsin. O.U.S.: None
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K2119978 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler, SureForm 30 Reloads
Intuitive Surgical, Inc., Dec 10, 2021